Model 120B Section V
Paragraphs 5-57 to 5-65
5-57. HORIZONTAL DC SHIFT CHECK. Check dc b. Set vertical AC-DC to AC.
shift follows: as c. Turn vertical VERNIER to CAL.
a: Set HORIZONTAL SENSITIVITY to .1 VOLTS/ d. Set HORIZONTAL DISPLAY to 5 pSEC/CM.
CM. e. Set TRIGGER SOURCE to EXT.
b. Turn horizontal VERNIER to CAL. f . Set TRIGGER LEVEL to AUTO.
c. Short horizontal input terminals. g. Using oscillator, apply a sine wave to both
d. Center spot and alternately switch horizontal the vertical and horizontal input terminals.
AC-DC to AC and DC positions. h. Vary oscillator frequency from 50 cps to 450
e. Spot should shift less than 0.2 cm. kc, maintaining a 0.53-volt input (1.5 volts peak-to-
peak).
5-58. HORIZONTAL ATTENUATOR-ERROR CHECK. i. Presentation should be stable over frequency
Check attenuator error as follows: range in step h.
a. Set HORIZONTAL DISPLAY to .1 VOLTS/CM.
b. Turn horizontal VERNIER to CAL. 5-62. SYNC - POLARITY AND TRIGGER- LEVEL-
RANGE CHECK. Check syne polarity and trigger-
c. Using voltmeter calibrator, apply a 400-cps level range as follows:
1-volt peak-to-peak sine wave to the horizontal input
terminals. a. Set VERTICAL SENSITIVITY to 10 VOLTS/CM.
b. Turn vertical VERNIER to CAL.
d. Adjust horizontal gain potentiometer R104 (fig- c. Set TRIGGER SOURCE to EXT.
ure 5-1) to give 10-cm deflection.
e. Change HORIZONTAL DISPLAY to 1 VOLTS/ d. Apply a 7.0-volt r-m-s 400-cps signal to ver-
CM and voltmeter calibrator output to 10 volts peak- tical and external sync inputs.
e. Center pattern on crt.
to-peak. f . Rotate TRIGGER LEVEL; triggering should be
f . Horizontal deflection should be &0.5 10 cm.
g. Change HORIZONTALSENSITIVITY to 10 VOLTS/ possible at any point along the negative slope of the
input signal.
CM and voltmeter calibrator output to 100 volts peak-
to-peak. 5-63. SWEEP-TIME CHECK. Check the sweep times
h. Horizontal deflection should be 10 *0.5 cm. as follows:
a. Set VERTICAL SENSITMTY to 1 VOLT/CM.
5-59. HORIZONTAL PHASE-SHIFT CHECK. Check b. Turn vertical VERNIER to CAL.
phase shift as follows: c. Set horizontal VERNIER to CAL.
a. Set VERTICAL SENSITIVITY to 100 MILLI- d. Set SWEEP MAGNIFIER to XI.
VOLTS/CM. e. Set TRIGGER SOURCE to INT-.
b. Turn vertical VERNIER to CAL. f. Adjust TRIGGER LEVEL for best synchroniza-
c. Set HORIZONTAL DISPLAY to .1 VOLTS/CM. tion.
d. Turn horizontal VERNIER to CAL. g. Connect time mark generator to vertical input
e. Set vertical and horizontal AC-DC to AC. terminals.
f. Using oscillator, apply a 100-kc 0.5-volt h. Referring to table 5-6, adjust the time mark
peak-to-peak sine wave to both vertical and horizontal generator output and the HORIZONTAL DISPLAY
input terminals. switch to the indicated markers and setting, respec-
g. Pattern opening on crt should be less than 0.2 tively, for each step; check for the indicated markers
cm (+2 degrees). per 10 cm listed in the table.
h. Change VERTICAL SENSITIVITY to 1 VOLTS/
CM, HORIZONTAL DISPLAY to 1 VOLTS/CM, and 5-64. CALIBRATOR AMPLITUDE CHECK. Check
voltmeter calibrator output to 5 volts peak-to-peak. the amplitude of the calibrator as follows:
i. Indication should be same as step g. a. Set VERTICAL SENSITIVITY to CAL.
j . Change VERTICAL SENSITMTY to 10 VOLTS/ b. Turn vertical VERNIER to CAL.
CM, HORIZONTAL DISPLAY to 10 VOLTS/CM, and c. Set HORIZONTAL DISPLAY to 1 MILLISEC-
voltmeter calibrator output to 50 volts peak-to-peak. OND/CM.
k. Indication should be same as step g. d. Set TRIGGER SOURCE to INT-.
e. Turn TRIGGER LEVEL to AUTO.
5-60. HORIZONTAL RESPONSE CHECK. Checkfre- f . Calibrator amplitude should be 6 *0.12 cm.
quency response as follows:
a. Set HORIZONTAL DISPLAY to 1 VOLTS/CM. 5-65. INTENSITY MODULATION CHECK. Check
b. Turn horizontal VERNIER to CAL. intensity modulation as follows:
c. Using oscillator, apply a 4.5-kc signal. to a. Set VERTICAL SENSITIVITY to 10 VOLTS/CM.
the horizontal input terminals. b. Turn vertical VERNIER to CAL.
d. Set oscillator output to provide a 10-cm trace. c. Set HORIZONTAL DISPLAY to 50 pSECONDS/
e. Monitor output with ac voltmeter and note voltage CM.
reading. d. Using ,square-wave generator, apply a 50-kc
f. ~aintaining voltage reading in step e, change 20-volt peak-to-peak square wave to the vertical in-
oscillator frequency to 300 kc. put terminals.
g; Trace length should be greater than 7 cm (3-db e. Remove shorting bar from Z axis terminal and
point). connect clip lead from vertical input to Z axis ter-
minal.
5-61. EXTERNAL TRIGGER-SENSITIVITY CHECK. f. With normal intensity, the top of square edge
Check external trigger sensitivity as follows: wave shouI& be extinguished when 20-volt peak-to-
a. Set VERTICAL SENSITIVITY to 1 VOLTS/CM. peak square wave is applied at the Z axis terminal. |