Model 175A Section V
Paragraphs 5-18 to 5-20
Table 5-2. Sweep Calibration
~~ ~
SWEEP TIME Markers Interval SWEEP TIME Markers Interval
.I ~SEC/CM 10 mc 1 cycle/cm 1 MSEC/CM 1 msec 1 marker/cm
.2 FSEC/CM 5 mc 1 cycle/cm 2 MSEC/CM 1 msec 2 markers/cm
.5 ~SEC/CM 1 psec 1 marker/2- cm 5 MSEC/CM 5 msec 1 marker/cm
1 ~SEC/CM 1 psec 1 marker/cm io MSEC/CM 10 msec 1 marker/cm
2 ~SEC/CM 1 psec 2 markers/cm 20 MSEC/CM 10 msec 2 markers/cm
5 FSEC/CM 5 psec 1 marker/cm 50 MSEC/CM 50 msec 1 marker/cm
io ~SEC/CM 10 psec 1 marker/cm .1 SEC/CM 100 msec 1 marker/cm
20 ~SEC/CM 10 psec 2 markers/cm .2 SEC/CM 100 msec 2 markers/cm
50 ~SEC/CM 50 psec 1 marker/cm .5 SEC/CM 500 msec 1 marker/cm
.I MSEC/CM 100 psec 1 marker/cm 1 SEC/CM 1 sec 1 marker/cm
.2 MSEC/CM 100 psec 2 markers/cm 2 SEC/CM 1 sec 2 markers/cm
.5 MSEC/CM 500 psec 1 marker/cm 5 SEC/CM 5 sec 1 marker/cm
w. A stable synchronized pattern should be obtained. d. The next marker should be within i0.5 cm of
right graticule edge.
x. Disconnect probe from power line and vertical
INPUT. e. Set:
SWEEP TIME ......... 0.1 pSEC/CM
5-18. SWEEP CALIBRATION.
a. Set: f. Set Time-Mark Generator output for 10 mc.
TRIGGER SOURCE. ........... INT g. Set one cycle of the sine wave to cross the left
SWEEP TIME. ........... 5 SEC/CM graticule edge.
Sweep VERNIER ... fully counterclockwise
SWEEP MODE .......... FREE RUN
h. Except for the first cycle of the expanded dis-
play, the corresponding point on the waveform one
b. It should take at least 125 seconds for the spot cycle later should cross within i0.5 cm of the right
to travel 10 cm. graticule edge.
c. Set:
SWEEP MODE ............ PRESET i. Disconnect Time-Mark Generator.
Sweep VERNIER ............. CAL
5-20. VERTICAL AMPLIFIER BANDWIDTH.
d. Connect the Time-Mark Generator to vertical a. Install Vertical Response Tester in place of
INPUT. Set Time-Mark Generator output and SWEEP vertical plug-in unit.
TIME as shown in table 5-2.
b. Set:
e. Adjust TRIGGER LEVEL and vertical SENSI-
SWEEP TIME. ......... 0.1 pSEC/CM
TIVITY, if necessary, for a stable, usable display.
c. Adjust HORIZONTAL POSITION and TRIGGER
f. In each case adjust HORIZONTAL POSITION LEVEL, if necessary, for a stable display.
to make the first marker coincide with theleft grati-
cule edge. The 10-cm marker should be within i0.3
d. Set Test Plug-In for a positive 5-cm pulse.
cm of the right graticule edge.
e. Center trace with VERTICAL POSITION. The
5-19. SWEEP MAGNIFIER. 10% and 90% points on the pulse rise time are now
a. Set: 2 cm below and 2 cm above the main horizontal axis
SWEEP TIME .......... 0.1 MSEC/CM of the crt.
HORIZONTAL DISPLAY ......... XI0
f. Set HORIZONTAL POSITION to place the 10%
point on any vertical line of the graticule. The 90%
b. Set Time-Mark Generator output for 100 psec point should be no more than 7 mm to the right of the
markers.
10% point,
c. With HORIZONTAL POSITION, set one marker g. Remove Test Plug-In and replace verticalplug-
to coincide with left graticule edge. in unit.
0 1526- 1 5-3 |